Skip to main content
ENעב
PRODUCTS
ENעב

Blog & Resources

Load Cell Calibration: Two-Point and Multi-Point Methods

Load cell calibration is the process of applying known forces to a load cell and recording its output so that every future reading can be converted into a trusted force value. A basic calibration uses two points, an unloaded zero and one known weight near capacity; a full calibration uses several increasing and decreasing load steps and fits a curve through them. Either way, the result is only as good as the reference loads and the care taken during the procedure.

Key takeaways

  • A two-point calibration needs a zero reading and one known weight; the gain follows from Weight = (Voltage Ratio − Offset) × Gain.
  • Use a reference load close to the load cell capacity, but never above it, because overload can permanently shift the zero.
  • Load cells typically output 1 to 3 mV/V at rated load, so a 2.0 mV/V cell on 5 V excitation gives 10 mV at full scale.
  • Multi-point and curve-fit calibration (the NIST handbook uses a quadratic fit) reveals nonlinearity that two points hide.
  • Traceable force calibration, for example to ISO 376, is done at a laboratory and documented with a certificate.

What Does Load Cell Calibration Actually Do?

Load cell calibration establishes the relationship between the force applied to the cell and the electrical signal it returns. A load cell converts force into a small voltage ratio using strain gauges bonded to a metal spring element, typically 1 to 3 mV/V at rated load. That raw ratio means nothing to an operator until calibration assigns it a unit: newtons, kilograms or pounds.

Calibration fixes two things. The first is the offset, the reading with no load, which links to the load cell zero balance specification. The second is the gain, the number of force units per unit of signal. Together they turn a voltage ratio into a measurement. If either one is wrong, every reading the system ever shows is wrong by the same systematic amount.

Calibration is not the same as accuracy

Calibration corrects the average relationship, but it cannot remove the cell’s inherent nonlinearity, hysteresis or repeatability limits. Those are design properties explained in our guide to load cell accuracy, nonlinearity and hysteresis. A well-calibrated cell with poor hysteresis still reads differently when loading and unloading; a precision cell with a bad calibration still reads the wrong value.

How Do You Calibrate a Load Cell With Two Points?

The two-point method is the simplest approach and is described, for example, in the Phidgets load cell documentation. It needs the cell installed in its final mounting, connected to the amplifier or interface, and one known weight.

Step 1: record the zero offset

With nothing on the system other than its permanent structure, record the output. The goal is to determine the offset of the system and remove it from all later measurements. Because the offset includes the weight of platforms and fixtures, repeat this step whenever the mechanical setup changes.

Step 2: apply one known reference load

Apply a certified weight that is close to the system capacity without overloading it. The Phidgets guide gives the example of a 4.8 kg weight on a 5 kg scale. A larger reference load makes the gain calculation less sensitive to small reading errors, which is why a tiny test weight on a high-capacity cell gives a poor calibration.

Step 3: compute the gain

The calibrated reading follows the formula Weight = (Voltage Ratio − Offset) × Gain. With the offset known from step 1 and the reference weight known from step 2, the gain is the weight divided by the difference between the loaded and unloaded readings. The gain then stays constant, while the offset is re-tared during normal use.

Illustrative arithmetic, not a measured result: a 2.0 mV/V cell on 5 V excitation produces 10 mV at rated load. If a 4.8 kg reference on a 5 kg cell raises the reading by 9.6 mV, the gain is 0.5 kg per mV, and a later reading of 4.0 mV above the offset corresponds to 2.0 kg.

When Do You Need Multi-Point Calibration?

Two points define a straight line, so a two-point calibration assumes the cell is perfectly linear. Real cells deviate slightly between the endpoints, and some applications cannot tolerate that. Multi-point calibration applies several known loads across the range, usually in increasing and then decreasing steps, and records the output at each.

Fitting a curve through the points

The NIST/SEMATECH Engineering Statistics Handbook (section 4.6.1, Load Cell Calibration) shows a quadratic calibration: a regression model relates a known load to the deflection of the cell, and the fitted model is then used to calibrate future readings associated with loads of unknown magnitude. The handbook walks through model fitting, residual analysis and model refinement, which is a useful template for any lab that wants a documented fitting procedure.

What the extra points reveal

The increasing and decreasing sequence exposes hysteresis, and the spread between repeated runs exposes repeatability. If the readings at the same load differ depending on the direction of approach, a straight-line gain cannot be correct in both directions, and the system should either use a curve or accept a stated tolerance.

Calibration Methods Compared

Method Reference loads What it corrects Typical use
Zero only (tare) None Offset Routine re-zeroing between uses
Two-point Zero plus one known weight Offset and gain Scales, test rigs, process monitoring
Multi-point Several steps, up and down Offset, gain and visible nonlinearity and hysteresis Test benches, R&D, quality labs
Curve fit (for example quadratic) Several steps, fitted by regression Systematic curvature in the response Precision measurement software
Traceable laboratory calibration Reference standard machine All of the above, documented with a certificate Regulated or audited measurement

What Goes Wrong During Calibration?

Most bad calibrations come from the setup, not the cell. The usual causes are:

  • Calibrating before installation is final: mounting hardware, cables and fixtures change the offset and the load path, so calibrate in the final configuration.
  • Exceeding capacity: an overload during calibration can permanently shift the zero balance, which defeats the purpose of the exercise.
  • Off-axis loading: a reference weight that is not applied along the sensing axis adds side forces that the cell reads as error.
  • Unstable reference weights: uncertified or contaminated weights put an error into the gain that no later correction can remove.
  • Skipping the re-tare: the offset drifts with temperature and handling, so it should be recorded again before each measurement series.

When Is Traceable Calibration Required?

An in-house check against certified weights is enough for many scales and monitoring systems. When results must stand up in an audit, in a regulated process or in a contract dispute, the calibration has to be traceable to a national standard and documented. ISO 376 is the standard widely used for calibrating force-proving instruments, as discussed in HBK’s article on the load cell calibration standard ISO 376. Laboratory calibration is performed against a reference force machine and returns a certificate listing the measured points.

Many manufacturers calibrate before shipping. FUTEK, whose range NewRoad supplies in Israel as authorized distributor since 2006, operates as a certified calibration laboratory and can supply its products with a calibration certificate. If your application needs documented force traceability, specify the certificate when you order, and check the options on the FUTEK company page, including the S-Beam load cell and the Universal Pancake for compression and tension. For construction types and how to pick between them, see our overview of the strain gauge load cell.

Frequently Asked Questions

What is load cell calibration?

It is the process of applying known forces to a load cell and recording its output, so that the output can be converted into a force value. It fixes the zero offset and the gain of the measurement chain.

How do you calibrate a load cell with a known weight?

Record the unloaded output as the offset, apply a certified weight close to capacity without exceeding it, and calculate the gain so that Weight equals (Voltage Ratio minus Offset) times Gain.

How many points does a load cell calibration need?

Two points (zero and one span load) are enough for simple systems. Multi-point calibration with increasing and decreasing steps is used when nonlinearity and hysteresis matter.

Is calibration the same as zero balance?

No. Zero balance is the output at no load, a single point. Calibration relates output to applied force across the measurement range, and zero is only its starting point.

What is ISO 376?

ISO 376 is an international standard for the calibration of force-proving instruments such as reference load cells. Laboratories use it to produce traceable calibration certificates.

Where can I get calibrated FUTEK load cells in Israel?

NewRoad is the authorized FUTEK distributor in Israel and supplies the range with a calibration certificate and local support.

Need a calibrated force sensor for your test setup?

Talk to NewRoad

Browse the full mechanical sensors and load cell technology page for specifications by sensor type.

Sources: Phidgets, Calibrating Load Cells; NIST/SEMATECH e-Handbook of Statistical Methods, 4.6.1 Load Cell Calibration; HBK, The Load Cell Calibration Standard ISO 376 (verified October 2026). The 2.0 mV/V arithmetic example is illustrative, not a measured value.

Last updated: October 2026.

Need calibrated FUTEK load cells for your lab?

Contact NewRoad

Share Post
More Posts
Atomic force microscope cantilever above a soft sample during a force curve measurement in liquid

AFM Force Spectroscopy: Force Curves, Stiffness and Adhesion

November 22, 2026
• 8 min read
Compact atomic magnetometer probe with optical fiber connections on a research bench

Atomic Magnetometer: How It Works and Where It Is Used

November 22, 2026
• 7 min read
Floor-standing scanning electron microscope with a sample chamber and display in an electron microscopy laboratory

Field Emission SEM vs Tungsten SEM: Which Do You Need?

November 22, 2026
• 7 min read
Atomic force microscope scan head with cantilever and camera view above a sample stage

How Does an Atomic Force Microscope Work? A Plain Guide

November 22, 2026
• 8 min read
Optical tensiometer camera imaging a liquid drop on a flat sample for contact angle measurement

Contact Angle Measurement: Methods, Meaning and Pitfalls

November 22, 2026
• 8 min read
Confocal Raman microscope with a laser line and motorized sample stage in a clean laboratory

Raman Microscope: How It Works and What It Can Do

November 22, 2026
• 9 min read